The LakeShore 335 Temperature Controller has been programmed by Bruker with default settings applicable to Dimension Icon Electrochemical Cell applications. Some of these are modifications of LakeShore’s original defaults. Both settings are listed below.
Figure 1: Controller Default Settings Reset Prompt Screen
Figure 2: Second Controller Default Settings Screen
Only the parameters appearing in bold font in the “Bruker Default” column have values different from the LakeShore default, so need to be changed. To reset individual parameters to the Bruker default:
Temperature Controller Button |
Temperature Controller Parameter |
Bruker Default | LakeShore Default |
---|---|---|---|
Display Setup | Two Input - Loop B | Two Input - Loop A | |
Output Setup |
Output Mode |
Closed Loop PID |
Closed Loop PID |
Control Input |
Input B |
Input A |
|
Power Up |
Disable |
Disable |
|
Heater Output Display |
Current |
Current |
|
Setpoint Ramp |
Off |
Off |
|
Input Setup - Inputs A & B: |
Sensor Type |
NTC RTD |
Diode |
Curve |
21 User 2252 |
DT-670 |
|
Current Reversal |
On |
On |
|
PID |
Proportional (P) |
50 |
50 |
Integral (I) |
20 |
20 |
|
Derivative (D) |
1 |
0.000 |
|
Curve Entry: |
(Curve 21 already installed) |
21 User 2252 |
N/A |